Endress+Hausert-mass

Proline t-mass A 150

Thermal Mass Flowmetert-mass - 6AAB

Flowmeter for measurement and easy monitoring of utility gases, such as air, nitrogen, carbon dioxide and argon in small line sizes.

Overview

Product Description

Proline t-mass A 150 is a thermal mass flowmeter for utility gases, in particular compressed air, and sub-metering applications. It offers 4-wire technology which is contained within a rugged compact aluminum housing. t-mass A 150 is available as a blind version or with display. Customer-specific settings are saved on the display and can be transferred to another device via the display.

Key Benefits

  • Suitable for air, nitrogen, carbon dioxide and argon in small line sizes
  • Optimal process monitoring – easy measurement even at low pressures and flow velocities
  • Easy installation, negligible pressure loss and maintenance-free
  • Fast and efficient commissioning thanks to guided operating menus
  • High plant availability due to self-diagnostics and error monitoring

Specifications

Measuring range
0.5 to 910 kg/h (1.1 to 2002 lb/h) 0.5 to 1365 kg/h (1.1 to 3003 lb/h) (for air, depending on chosen option of ordering feature "Calibration flow")
Max. measurement error
3 % o.r. 4 % o.r. 5 % o.f.s. (depending on chosen option of ordering feature "Calibration flow")
Wetted materials
Transducer: 1.4404 (316L) Insertion tube: 1.4404 (316L) Measuring tube: 1.4404 (316L); 1.4435 (316L) Connection: 1.4404 (F316/F316L); 1.4404 (316L); 1.4435 (316L)
Medium temperature range
–40 to +100 °C (–40 to +212 °F)
Max. process pressure
PN 40, Class 300

Measuring task

Product headline
The flowmeter for cost-effective measurement and easy monitoring of utility gases. Suitable for air, nitrogen, carbon dioxide and argon in small line sizes.
Sensor features
Optimal process monitoring – easy measurement even at low pressures and flow velocities. Cost-effective measurement – easy installation, negligible pressure loss and maintenance-free. Reliable flow trending – multivariable measurement. Inline version: DN 15 to 50 (½ to 2"). Process pressure up to PN 40, Class 300. A variety of process connections available.
Transmitter features
Fast and efficient commissioning – guided operating menus. High plant availability – self‐diagnostics and error monitoring. Automatic recovery of data for servicing. Device in compact version with DC 24 V power supply. 4-20 mA HART, pulse/frequency/switch output. Compact and robust transmitter.
Nominal diameter range
DN 15 to 50 (½ to 2")
Wetted materials
Transducer: 1.4404 (316L) Insertion tube: 1.4404 (316L) Measuring tube: 1.4404 (316L); 1.4435 (316L) Connection: 1.4404 (F316/F316L); 1.4404 (316L); 1.4435 (316L)
Measured variables
Mass flow, temperature, corrected volume flow, FAD volume flow
Max. measurement error
3 % o.r. 4 % o.r. 5 % o.f.s. (depending on chosen option of ordering feature "Calibration flow")
Measuring range
0.5 to 910 kg/h (1.1 to 2002 lb/h) 0.5 to 1365 kg/h (1.1 to 3003 lb/h) (for air, depending on chosen option of ordering feature "Calibration flow")
Max. process pressure
PN 40, Class 300
Medium temperature range
–40 to +100 °C (–40 to +212 °F)
Ambient temperature range
–40 to +60 °C (–40 to +140 °F)
Transmitter housing material
AlSi10Mg, coated
Degree of protection
IP66/67, type 4X enclosure
Display/Operation
4‐line display with push Buttons Configuration via local display and operating tools possible
Outputs
4‐20 mA HART (active) Pulse/frequency/switch output (passive)
Inputs
Status input
Digital communication
HART
Power supply
DC 18 to 30 V
Hazardous area approvals
ATEX, IECEx, cCSAus
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR
Pressure approvals and certificates
PED, CRN

Technical documents

Operational documents

Videos

Industries:Mining, Minerals & MetalsPower & energyWater & wastewater

Keywords:tmassthermal flowmeterflow metermass flowutility