Proline t-mass I 500
Insertion flowmeter with long-term stability as remote version with up to 4 I/Os. Measurement of utility and process gases and gas mixtures in circular piping or rectangular ducts.
Overview
Specifications
- Max. measurement error
- Gas: 1.0% o.r. (10 to 100% o.f.s.), 0.1% o.f.s. (1 to 10% o.f.s.)
- Medium temperature range
- -40 °C to +180°C (-40 °F to +356°F)
- Wetted materials
- Materials for insertion tube Stainless steel, 1.4404 (316/316L) Process connections, process coupling Stainless steel, 1.4404 (316/316L) Sensing element Unidirectional Stainless steel, 1.4404 (316/316L) Alloy C22, 2.4602 (UNS N06022); Bidirectional Stainless steel, 1.4404 (316/316L) Reverse flow detection Stainless steel, 1.4404 (316/316L) Clamping rings PEEK PVDF 1.4404 (316/316L) Flat ring seal EPDM FKM
- Max. process pressure
- -0.5 to 20 bar_g (-7.25 to 290 psi_g)
- Measuring range
- 20 to 733501 kg/h (44 to 1669340 lb/h)
Measuring task
- Product headline
- Insertion flowmeter with long-term stability as remote version with up to 4 I/Os. Flexible, convenient programming based on 21 standard gases or freely definable gas mixtures thereof. Measurement of utility and process gases as well as gas mixtures in circular piping or rectangular ducts.
- Sensor features
- High level of process control – premium measurement accuracy and repeatability. Reliable monitoring – detection of process disturbances and reverse flow. Flexible installation – suitable for large dimensional range and circular pipes or rectangular ducts. Insertion version for DN 80 to 1500 (3 to 60"). Bidirectional measurement; high measuring performance. Patented drift-free sensor with SIL 2.
- Transmitter features
- Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology. Remote version with up to 4 I/Os. Backlit display with touch control and WLAN access. Standard cable between sensor and transmitter.
- Nominal diameter range
- DN 80 to 1500 (3 to 60")
- Wetted materials
- Materials for insertion tube Stainless steel, 1.4404 (316/316L) Process connections, process coupling Stainless steel, 1.4404 (316/316L) Sensing element Unidirectional Stainless steel, 1.4404 (316/316L) Alloy C22, 2.4602 (UNS N06022); Bidirectional Stainless steel, 1.4404 (316/316L) Reverse flow detection Stainless steel, 1.4404 (316/316L) Clamping rings PEEK PVDF 1.4404 (316/316L) Flat ring seal EPDM FKM
- Measured variables
- Massflow, temperature, standard volume flow, volume flow, Free air delivery, velocity, heat flow, energy flow, density
- Max. measurement error
- Gas: 1.0% o.r. (10 to 100% o.f.s.), 0.1% o.f.s. (1 to 10% o.f.s.)
- Measuring range
- 20 to 733501 kg/h (44 to 1669340 lb/h)
- Max. process pressure
- -0.5 to 20 bar_g (-7.25 to 290 psi_g)
- Medium temperature range
- -40 °C to +180°C (-40 °F to +356°F)
- Ambient temperature range
- -40 to 60°C (-40 to 140°F) Optional: Transmitter: -50 to 60°C (-50 to 140°F), Sensor: -60 to 60°C (-60 to 140°F)
- Transmitter housing material
- Aluminium, AlSi10Mg, coated Polycarbonate
- Degree of protection
- IP66/67, Type 4X enclosure Sensor: IP68, Type 6P (optional)
- Display/Operation
- 4-line backlit display with touch control (operation from outside) Configuration via local display and operating tools possible Remote display available
- Outputs
- 4-20 mA HART (active/passive) 4-20 mA (active/passive) Pulse/frequency/switch output (active/passive) Relay output
- Inputs
- Status input 4-20 mA input
- Digital communication
- HART, Modbus RS485
- Power supply
- DC 24V AC 100 to 240V
- Hazardous area approvals
- ATEX, cCSAus, IECEx, NEPSI, JPN, UK Ex, EAC, KC
- Product safety
- CE, C-tick
- Functional safety
- Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
- Metrological approvals and certificates
- Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025) Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a
- Pressure approvals and certificates
- CRN
- Material certificates
- 3.1 material NACE MR0175/MR0103
Technical documents
Operational documents
Related products
t-mass - 65I